DTC optimization: Difference between revisions
(9 intermediate revisions by the same user not shown) | |||
Line 12: | Line 12: | ||
== Results == | == Results == | ||
=== Resolution === | |||
All following values are given in units of Water Equivalent Thickness! (Except the absorber thicknesses) | All following values are given in units of Water Equivalent Thickness! (Except the absorber thicknesses) | ||
Line 28: | Line 31: | ||
| Al || 4 mm || 2.7 mm || 4.35 mm || 12.4% || 1.97 mm || 39 | | Al || 4 mm || 2.7 mm || 4.35 mm || 12.4% || 1.97 mm || 39 | ||
|- | |- | ||
| Al || 5 mm || 3.3 mm || | | Al || 5 mm || 3.3 mm || 4.50 mm || 16.4% || 2.28 mm || 32 | ||
|- | |||
| Al || 6 mm || 3.9 mm || 4.70 mm || 23.3% || 2.73 mm || 27 | |||
|- | |- | ||
| Al || | | Al || 7 mm || 4.5 mm || 4.76 mm || 26.0% || 2.78 mm || 23 | ||
|- | |- | ||
| | | Loma Linda || N/A || N/A || 4.1 mm* || N/A || N/A || N/A | ||
|- | |||
| FOCAL || 32 mm || 9.2 mm || 15.6 mm** || N/A || N/A || 11 | |||
|} | |} | ||
* At 200 MeV. Scintillator measurements, res. does not scale linearly with depth. | |||
** Resolution scaled to 230 MeV from 188 MeV | |||
[[File:resolution_2mm.png|1200px]] | [[File:resolution_2mm.png|1200px]] | ||
Line 41: | Line 50: | ||
[[File:resolution_4mm.png|1200px]] | [[File:resolution_4mm.png|1200px]] | ||
[[File:resolution_5mm.png|1200px]] | |||
[[File:resolution_6mm.png|1200px]] | |||
[[File:resolution_7mm.png|1200px]] | |||
=== Accuracy === | |||
[[File:accuracy_2mm.png|800px]] | |||
[[File:accuracy_3mm.png|800px]] | |||
[[File:accuracy_4mm.png|800px]] | |||
[[File:accuracy_5mm.png|800px]] | |||
[[File:accuracy_6mm.png|800px]] | |||
[[File:accuracy_7mm.png|800px]] |
Latest revision as of 09:17, 30 March 2017
Geometry and segmentation optimization of the Digital Tracking Calorimeter
Optimization philosophy
Geometry
Materials
Software
The software for this project is Gate installation and the DTC toolkit. For a complete guide to use the software for optimization, see the page Software for design optimization.
Results
Resolution
All following values are given in units of Water Equivalent Thickness! (Except the absorber thicknesses)
Pre-straggling resolution is the contribution to the total resolution from the detector smearing. This is calculable because we know the range straggling from Monte Carlo simulations, and the pre-straggling resolution is a subtraction in quadrature.
Material | Thickness | [math]\displaystyle{ t / \sqrt{12} }[/math] WEPL | Resolution | Excess straggling | Pre-straggling resolution | Layers for 230 MeV + [math]\displaystyle{ 5\sigma }[/math] |
---|---|---|---|---|---|---|
Al | 2 mm | 1.5 mm | 4.16 mm | 6.1% | 1.38 mm | 67 |
Al | 3 mm | 2.1 mm | 4.26 mm | 9.2% | 1.69 mm | 48 |
Al | 4 mm | 2.7 mm | 4.35 mm | 12.4% | 1.97 mm | 39 |
Al | 5 mm | 3.3 mm | 4.50 mm | 16.4% | 2.28 mm | 32 |
Al | 6 mm | 3.9 mm | 4.70 mm | 23.3% | 2.73 mm | 27 |
Al | 7 mm | 4.5 mm | 4.76 mm | 26.0% | 2.78 mm | 23 |
Loma Linda | N/A | N/A | 4.1 mm* | N/A | N/A | N/A |
FOCAL | 32 mm | 9.2 mm | 15.6 mm** | N/A | N/A | 11 |
* At 200 MeV. Scintillator measurements, res. does not scale linearly with depth. ** Resolution scaled to 230 MeV from 188 MeV